Glennan Room 308   LAB NOTES

 

--R. A. Leskovec   January 30, 2001

 

 

 

There are some problems that can potentially severely limit the amount of working stations in the 308 lab and inconvenience yourself as well as everyone else.

 

Through carelessness and lack of focus, the users cause fuses to be blown in the DMM's (Digital MultiMeters) and in the Wave generator, which cannot readily be replaced.  This removes the units from service until I am able to repair them.  On the other hand, we have found, that once the situation is carefully and clearly explained, the incidents go practically to zero!   So  we appreciate everyone’s cooperation.

 

In the case of the DMM's,  please clearly understand the difference between measuring voltage and current, so that you do not hook the current function directly across a power source, as this will blow the fuse.  I can replace these with some difficulty, but at least do not have to move the table, to remove the cable that ties down all the equipment.

 

In the case of the HP33120 Function Generator, the problem is very serious because I have to completely remove the unit, and take it to the shop to replace an internal fuse that is soldered in!  So please be cautious to carefully check your circuits and NEVER connect a voltage source such that it feeds back into the OUTPUT of the generator!  This includes power supplies as well as another generator, even if the subject  generator is OFF!

 

Incidentally, scope probes do not have any decent bandwidth in the x1 mode.  They have more bandwidth in the x10 mode but need to be calibrated.   The scopes have calibration signals on the front panels, which also give you a way of checking that both the probe and scale factor of the scope are correct.  Calibration procedures are explained on a large HP "poster" to the right of the chalk board in the front of the room, and the little “tools” are in the stockroom tray.

 

If you suspect there is a problem with your generator or DMM, the Lab Assistant in the stockroom has a loose-leaf  “Operations and Information Manual” in which there are some troubleshooting sheets that may help you diagnose the problem.  There are also some labels on the instruments themselves.  But if it things still don’t work, please send email including the table number to:

 

help@eecs.cwru.edu